Care Max: Arduino Heart Rate Monitoring System #
Designing a real-time cardiovascular telemetry system with Arduino microcontrollers, photoplethysmography sensors, and embedded C++.
Demonstration Notice: This article documents an academic embedded biomedical engineering project, including hardware schematics, firmware algorithms, and laboratory signal benchmarking to illustrate the engineering scope and implementation approach.
Overview #
Cardiovascular disease remains the leading cause of global mortality. While modern hospital telemetry systems and high-end consumer smartwatches monitor heart rates effectively, there is a persistent need for low-cost, transparent, and accessible biomedical monitoring hardware for educational laboratories, rural clinics, and student biomedical research.
I developed Care Max as an open, accessible Arduino-based cardiovascular monitoring system. Utilizing photoplethysmography (PPG) optical sensing, Care Max measures microscopic changes in capillary blood volume, calculates real-time Beats Per Minute (BPM), and renders live pulse waveform graphs and physiological metrics on hardware OLED displays and serial telemetry terminals.
Problem and Context #
Processing biological signals on resource-constrained 8-bit microcontrollers presents intense hardware and software challenges:
- Severe Motion Artifacts: Minor finger shifts, sensor pressure variations, and ambient optical noise (such as room lighting) distort subtle blood volume pulses.
- Limited Memory and CPU Capacity: Microcontrollers like the ATmega328P possess only 2 KB of SRAM and run at 16 MHz, precluding heavy floating-point digital signal processing libraries.
- Real-Time Responsiveness: The system must capture analog sensor samples at a strict, deterministic sampling rate while simultaneously driving serial telemetry and screen refreshes without dropped pulses.
Objectives and Requirements #
- Accurate BPM Calculation: Compute resting heart rates within +/- 2 BPM of commercial clinical pulse oximeters.
- Hardware-Constrained Signal Filtering: Implement lightweight digital low-pass and peak-detection algorithms that execute efficiently within 2 KB of SRAM.
- Real-Time Waveform Display: Render continuous pulse waveforms and status indicators on an I2C OLED display at 30 updates per second.
- Audible and Visual Pulse Cues: Synchronize a piezoelectric buzzer tone and onboard LED blink with each detected ventricular contraction (heartbeat).
Approach and Process #
I built Care Max using focused embedded systems components and low-level firmware engineering:
- Microcontroller Hardware: Built on the Arduino Uno (ATmega328P) operating at 5V, 16 MHz.
- Sensor Module: Used a Pulse Sensor Amped (optical green LED photoplethysmography sensor with integrated operational amplifier).
- Display Output: Interfaced a 0.96-inch SSD1306 monochrome OLED (128x64 resolution) communicating via hardware I2C bus.
- Programming Language & Libraries: Wrote Embedded C++ utilizing timer interrupts (
Timer2) for deterministic sampling, coupled with the Adafruit GFX library for lightweight display drawing. - Development Tooling: Programmed via Arduino IDE, AVR-GCC toolchain, and tracked firmware with Git/GitHub.
Solution and Key Features #
The embedded firmware runs an interrupt-driven sampling architecture to decouple data acquisition from display rendering:
[Hardware Timer2 (2ms Interrupt)] ──► Fires Every 2,000 Microseconds (500 Hz)
│
├─► Read ADC Analog Pin (Sensor Voltage)
├─► Apply Rolling Average Filter
├─► Peak / Trough Detection & Threshold Crossing
└─► Calculate Inter-Beat Interval (IBI) & BPM
│
▼
[Volatile Memory Buffers]
│
▼
[Main Execution Loop]
├─► Draw Pulse Waveform on OLED Display (I2C)
├─► Trigger Piezo Tone on Heartbeat
└─► Stream Serial Telemetry ("BPM: 72, IBI: 833ms")
Deterministic Interrupt Sampling: Rather than relying on
delay()functions in the main loop, I configured an internal hardware timer interrupt (Timer2) to trigger exactly every 2 milliseconds (500 Hz). This guarantees an unjittered time-series sampling of the arterial pulse wave.Adaptive Peak Detection Algorithm: The firmware dynamically tracks the signal's maximum peak and minimum trough values over a rolling 10-beat window. A heartbeat is registered when the signal crosses an adaptive threshold (50% of the peak-to-trough amplitude) during an ascending slope, provided at least 300 ms have elapsed since the prior beat (preventing false double-triggering on dichrotic notches): $$\text{BPM} = \frac{60{,}000}{\text{Inter-Beat Interval (ms)}}$$
Smooth Waveform Scrolling: The main loop maintains a rolling 128-byte array representing recent pulse heights. Each iteration shifts the screen coordinates by 1 pixel, drawing continuous line segments on the OLED display to produce a clean, continuous rhythm strip visual.
My Role and Contributions #
As the embedded systems and biomedical hardware integrator, my personal contributions included:
- Circuit Schematic & Prototyping: Breadboarded the complete circuit connecting the PPG sensor, I2C pull-up resistors, piezoelectric buzzer, and SSD1306 OLED display.
- Interrupt Service Routine (ISR) Architecture: Wrote the low-level AVR register configuration (
TCCR2A,TCCR2B,TIMSK2) to configure Timer2 CTC mode at a 500 Hz interrupt rate. - Signal Filtering & Beat Detection: Developed the digital moving-average filtering algorithm and dynamic threshold tracking logic in C++.
- Display Driver Optimization: Tuned the OLED rendering loop, drawing vertical raster lines rather than re-clearing the entire display buffer to conserve execution cycles.
- Laboratory Calibration: Tested and benchmarked sensor readings against a commercial finger-clip pulse oximeter across resting subjects.
Challenges and Decisions #
- Mains Electricity 60 Hz Ambient Hum:
- Challenge: Fluorescent lighting in test rooms induced 60 Hz optical noise spikes on the phototransistor analog input.
- Decision: I implemented a software moving-average low-pass filter in C++ that smooths high-frequency spikes prior to peak evaluation.
- I2C Bus Bottlenecks:
- Challenge: Updating the full 128x64 OLED display buffer over standard 100 kHz I2C slowed down the main loop.
- Decision: I increased the I2C clock frequency to Fast Mode (400 kHz) via
TWBR = 12, quadrupling screen refresh rates.
- Memory Footprint Constraints:
- Challenge: The ATmega328P has only 2048 bytes of SRAM; graphic buffers and arrays risked causing stack-heap collisions.
- Decision: I stored static glyphs and strings in Flash memory using the
PROGMEMandF()macros, keeping dynamic RAM usage under 1.1 KB.
Results and Validation #
- Accuracy Verification: Measured resting heart rates within +/- 1.8 BPM of reference commercial finger oximeters in laboratory test trials.
- Sampling Stability: Maintained a stable 500 Hz sampling cadence with zero missed interrupt ticks during continuous 60-minute monitoring trials.
- Resource Footprint: The entire firmware consumed only 14.2 KB of Flash (46%) and 1.1 KB of SRAM (55%), leaving ample headroom for future wireless Bluetooth expansions.
- Real-Time Visuals: Achieved fluid 30 FPS waveform scrolling on the OLED display with synchronized audio beat clicks.
Lessons Learned and Future Improvements #
- Interrupts are Critical for Signal Processing: Keeping analog sampling inside dedicated hardware interrupts prevents UI rendering operations from distorting real-time sensor data.
- Accessible Embedded Tools Have High Educational Value: Simple, low-cost microcontrollers can deliver genuine biomedical insight when paired with sound mathematical filtering and careful C++ memory management.
- Future Roadmap: I plan to port the firmware to an ESP32 microcontroller with BLE GATT profiles, allowing heart rate data to stream directly to mobile health applications.
Closing and Project Links #
Care Max demonstrated that rigorous interrupt timing and adaptive digital filtering can transform accessible microcontroller hardware into a reliable, real-time biomedical monitoring tool.